Code on display Err Lo RADWAG WTC 600

Single piece weight too small / min > max

In piece counting mode, the determined single piece weight is less than 0.1 of the reading unit, or in +/- checkweighing mode the Min threshold was entered greater than the Max threshold.

You can usually fix this yourself.

What should you do?

Try the steps in order.

  1. In piece counting, use more reference pieces.
  2. In +/- checkweighing mode, check and correct the Min and Max threshold values.

Possible causes

  • The number of reference pieces is low
  • Pieces too light
  • Min/Max thresholds entered in reverse

Do you need service?

The message clears once the settings or reference quantity are corrected. No service is needed.

Models showing this code: WTC 600, WTC

Source: RADWAG official user manual · Open manufacturer document