Code on display Err 8.4 OHAUS Navigator NV

Underload

The weight fell below the lowest measuring range while the balance was running. The manual advises re-installing the pan.

Check it yourself first. If it continues, it needs service.

What should you do?

Try the steps in order.

  1. Remove the pan and install it again correctly.
  2. Check that there is no foreign object under the pan.
  3. Zero the balance.

Possible causes

  • Pan removed or shifted
  • Pan installed incorrectly

Do you need service?

If Err 8.4 remains with the pan correctly installed, the load cell must be checked by a technician.

Models showing this code: NV222, NV622, NV1202, NV2202, NV3202

Source: OHAUS official user manual · Open manufacturer document