Code on display Err 3.0 OHAUS Navigator NV

Wrong adjustment weight

The balance does not accept the weight on the pan as a calibration point. Calibration cannot be completed with this weight.

Check it yourself first. If it continues, it needs service.

What should you do?

Try the steps in order.

  1. Use the calibration weight listed for your model in the manual.
  2. Move the scale to a place free of vibration and drafts.
  3. Restart calibration with an empty pan.

Possible causes

  • Calibration weight not suited to the model
  • Vibration or air currents during calibration

Do you need service?

If Err 3.0 remains with the correct weight, the balance electronics must be checked by a technician.

Models showing this code: NV222, NV622, NV1202, NV2202, NV3202

Source: OHAUS official user manual · Open manufacturer document