RS232 buffer full
The balance cannot pass its data to the connected device and pending data piles up. This happens with a printer or computer connection.
Check it yourself first. If it continues, it needs service.
What should you do?
Try the steps in order.
- Turn on RS232 handshake in the menu.
- Check that the printer or computer is on and ready to receive.
- Check that settings such as baud rate match on both devices.
- Apply the settings in the interface manual.
Possible causes
- Handshake setting is off
- Printer or computer is not receiving data
- Communication settings do not match
Do you need service?
If the buffer keeps filling with correct settings, the RS232 interface kit must be checked by a technician.
Models showing this code: NV222, NV622, NV1202, NV2202, NV3202
Source: OHAUS official user manual · Open manufacturer document
Legal notice
The error code explanations and checks on this page are based on manufacturer manuals and are provided for general information only; they do not replace technical service, diagnosis or warranty. The meaning of a code may vary by model, firmware version and configuration.
The decision to carry out any suggested step, and its consequences, rest with the user. Terazibul cannot be held liable for device damage, weighing errors, data loss, commercial loss or any other direct or indirect damage arising from the use of this information.
On legal-for-trade (verified) scales, adjustment and calibration must only be performed by a qualified technical service in line with legal metrology rules. Disconnect power before opening the device and contact a service provider if in doubt. Brand and model names belong to their respective owners and are used only to identify the device.