Code on display Err 8.1 OHAUS Navigator NV

Power-on zero range exceeded

The balance sees more weight than expected at start-up and cannot set its zero point. The manual also asks you to check the shipping lock.

Check it yourself first. If it continues, it needs service.

What should you do?

Try the steps in order.

  1. Empty the weighing pan.
  2. Set the shipping lock under the balance to unlocked.
  3. Check that nothing is touching the weighing pan.
  4. Switch the balance off and on again.

Possible causes

  • There was a load on the weighing pan at switch-on
  • Shipping lock in the wrong position
  • An object touching the pan

Do you need service?

If Err 8.1 remains with an empty pan and the lock released, the load cell must be checked by a technician.

Models showing this code: NV222, NV622, NV1202, NV2202, NV3202

Source: OHAUS official user manual · Open manufacturer document