Code on display Over initial zero range OHAUS MB92

Power-on zero range exceeded

The analyzer sees too much weight on the pan support at start-up and cannot set zero. Usually a sample or sample pan was left in place.

Check it yourself first. If it continues, it needs service.

What should you do?

Try the steps in order.

  1. Remove the sample and the sample pan from the pan support.
  2. Switch the device off and on again.
  • The heater and sample chamber may be very hot. Take care when opening the cover.

Possible causes

  • Sample left on the pan support at start-up
  • Sample pan left on the pan support
  • Load cell fault

Do you need service?

If Over initial zero range remains with an empty pan support, the load cell must be checked by a technician.

Models showing this code: MB92