Code on display Error 9.5 OHAUS Defender 5000 / D52

No calibration data

Without calibration data the indicator cannot calculate the weight accurately. The solution in the manual is to calibrate the scale.

Check it yourself first. If it continues, it needs service.

What should you do?

Try the steps in order.

  1. Check that the calibration menu lock and the LFT setting are off.
  2. Calibrate the scale with the correct calibration weight.
  • On devices in legal (LFT/approved) use, do not break the seal. The seal must be opened only by technical service.

Possible causes

  • Calibration not performed
  • Calibration data deleted

Do you need service?

On sealed devices, or if the calibration cannot be saved, technical service calibration is required.

Source: OHAUS official user manual · Open manufacturer document