Code on display Error 8.1 OHAUS Defender 5000 / D52

Power-on zero limit exceeded

The indicator cannot take the initial zero at power-up because there is more load on the platform than expected. The manual recommends removing the load and recalibrating if necessary.

Check it yourself first. If it continues, it needs service.

What should you do?

Try the steps in order.

  1. Remove all load from the platform.
  2. Turn the indicator off and on again.
  3. If the error continues, recalibrate the scale.

Possible causes

  • Load on the platform at power-up
  • Object resting against the platform
  • Calibration drift

Do you need service?

If the error continues with an empty platform and after calibration, contact technical service to check the load cell.

Source: OHAUS official user manual · Open manufacturer document