Code on display Error 8.4 OHAUS Defender 5000 / D52

Underload

The signal is below the allowed lower limit. It is mostly caused by the platform being lifted or jammed, or by calibration drift.

Check it yourself first. If it continues, it needs service.

What should you do?

Try the steps in order.

  1. Check that the platform is in place and free to move.
  2. If the error continues, recalibrate the scale.

Possible causes

  • Platform lifted or caught
  • Calibration drift

Do you need service?

If it continues despite calibration, contact service to check the load cell or cable.

Source: OHAUS official user manual · Open manufacturer document