Code on display Error 8.3 OHAUS Defender 5000 / D52

Overload

The load on the platform is above capacity. Loading beyond capacity can permanently damage the load cell.

You can usually fix this yourself.

What should you do?

Try the steps in order.

  1. Reduce the load on the platform.
  2. Weigh heavy loads in portions so they stay within capacity.
  • Do not drop the load onto the platform.

Possible causes

  • Load exceeding capacity
  • Shock loading

Do you need service?

If the error continues after the load is removed, contact service because the load cell may be damaged.

Source: OHAUS official user manual · Open manufacturer document