Base calibration data invalid
On bases connected with EasyConnect, the calibration data is stored in the memory module in the cable. If this data is not valid, the indicator cannot weigh correctly with the base. The manual recommends calibrating the scale.
Check it yourself first. If it continues, it needs service.
What should you do?
Try the steps in order.
- Calibrate the scale with the correct calibration weight.
- Turn the indicator off and on after calibration.
Possible causes
- Invalid calibration data in the memory module
- A different base connected
Do you need service?
If Error 8.8 continues after calibration, contact service to check the memory module or cable.
Source: OHAUS official user manual · Open manufacturer document
Legal notice
The error code explanations and checks on this page are based on manufacturer manuals and are provided for general information only; they do not replace technical service, diagnosis or warranty. The meaning of a code may vary by model, firmware version and configuration.
The decision to carry out any suggested step, and its consequences, rest with the user. Terazibul cannot be held liable for device damage, weighing errors, data loss, commercial loss or any other direct or indirect damage arising from the use of this information.
On legal-for-trade (verified) scales, adjustment and calibration must only be performed by a qualified technical service in line with legal metrology rules. Disconnect power before opening the device and contact a service provider if in doubt. Brand and model names belong to their respective owners and are used only to identify the device.