Code on display Error 8.8 OHAUS Defender 5000 / D52

Base calibration data invalid

On bases connected with EasyConnect, the calibration data is stored in the memory module in the cable. If this data is not valid, the indicator cannot weigh correctly with the base. The manual recommends calibrating the scale.

Check it yourself first. If it continues, it needs service.

What should you do?

Try the steps in order.

  1. Calibrate the scale with the correct calibration weight.
  2. Turn the indicator off and on after calibration.

Possible causes

  • Invalid calibration data in the memory module
  • A different base connected

Do you need service?

If Error 8.8 continues after calibration, contact service to check the memory module or cable.

Source: OHAUS official user manual · Open manufacturer document