Code on display CAL E OHAUS Defender 3000 / i-DT33

Calibration error

The calibration is not accepted because the value measured during calibration is not within the expected range. It is most often caused by using the wrong calibration weight.

You can usually fix this yourself.

What should you do?

Try the steps in order.

  1. Use the calibration weight that matches the value selected in the menu.
  2. Start the calibration with an empty, stable platform.
  3. Repeat the calibration in an environment free of vibration.

Possible causes

  • Wrong calibration weight
  • Unstable environment
  • Other load on the platform

Do you need service?

If the calibration is rejected even when repeated with the correct weights, contact service to check the load cell.

Source: OHAUS official user manual · Open manufacturer document