Code on display Error 8.4 OHAUS Defender 3000 / i-DT33

Underload

The signal is below the allowed lower limit of the zero point. It is usually caused by the platform being lifted or seated incorrectly, or by calibration drift.

Check it yourself first. If it continues, it needs service.

What should you do?

Try the steps in order.

  1. Check that the platform is in place and free to move.
  2. Zero the indicator or turn it off and on.
  3. If the error continues, recalibrate the scale.

Possible causes

  • Platform top removed
  • Platform caught on something
  • Calibration drift

Do you need service?

If Error 8.4 continues despite calibration, contact service for a possible load cell or cable fault.

Source: OHAUS official user manual · Open manufacturer document