Power-on zero limit exceeded
At power-up the indicator assumes the platform is empty and takes the initial zero. If a load remains on the platform or the calibration has drifted, zero cannot be taken. The manual recommends removing the load and recalibrating if necessary.
Check it yourself first. If it continues, it needs service.
What should you do?
Try the steps in order.
- Remove all load from the platform.
- Check that nothing is resting against the platform.
- Turn the indicator off and on again.
- If the error continues, recalibrate the scale with the correct calibration weight.
Possible causes
- Load on the platform at power-up
- Object resting against the platform
- Calibration drift
Do you need service?
If Error 8.1 continues with an empty platform and after calibration, contact technical service to check the load cell or connection.
Source: OHAUS official user manual · Open manufacturer document
Legal notice
The error code explanations and checks on this page are based on manufacturer manuals and are provided for general information only; they do not replace technical service, diagnosis or warranty. The meaning of a code may vary by model, firmware version and configuration.
The decision to carry out any suggested step, and its consequences, rest with the user. Terazibul cannot be held liable for device damage, weighing errors, data loss, commercial loss or any other direct or indirect damage arising from the use of this information.
On legal-for-trade (verified) scales, adjustment and calibration must only be performed by a qualified technical service in line with legal metrology rules. Disconnect power before opening the device and contact a service provider if in doubt. Brand and model names belong to their respective owners and are used only to identify the device.