Code on display Error 8.1 OHAUS Defender 3000 / i-DT33

Power-on zero limit exceeded

At power-up the indicator assumes the platform is empty and takes the initial zero. If a load remains on the platform or the calibration has drifted, zero cannot be taken. The manual recommends removing the load and recalibrating if necessary.

Check it yourself first. If it continues, it needs service.

What should you do?

Try the steps in order.

  1. Remove all load from the platform.
  2. Check that nothing is resting against the platform.
  3. Turn the indicator off and on again.
  4. If the error continues, recalibrate the scale with the correct calibration weight.

Possible causes

  • Load on the platform at power-up
  • Object resting against the platform
  • Calibration drift

Do you need service?

If Error 8.1 continues with an empty platform and after calibration, contact technical service to check the load cell or connection.

Source: OHAUS official user manual · Open manufacturer document