Code on display Error 8.2 OHAUS Defender 3000 / i-DT33

Below power-on zero limit

At power-up the indicator reads a signal lower than the expected dead load. This error appears if the weighing pan or platform top is missing or the calibration has drifted. The manual recommends adding load and recalibrating.

Check it yourself first. If it continues, it needs service.

What should you do?

Try the steps in order.

  1. Check that the platform top/weighing pan is in place and seated correctly.
  2. Turn the indicator off and on again.
  3. If the error continues, recalibrate the scale.

Possible causes

  • Platform top not installed
  • Platform seated incorrectly
  • Calibration drift

Do you need service?

If the error continues with the platform correctly installed and after calibration, contact service to check the load cell connection.

Source: OHAUS official user manual · Open manufacturer document