Code on display Error 9 Minebea Intec X3 PR 5410

No values from scale

At the internal weighing point, no value can be read from the A/D converter. Communication with the xBPI scale or digital load cells has been lost.

Check it yourself first. If it continues, it needs service.

What should you do?

Try the steps in order.

  1. Check that the connected scale or junction box is powered.
  2. Visually inspect the cables.
  3. Switch the device off and on again.
  • Do not open the device housing. Only technical service should work on internal settings and boards.

Possible causes

  • Broken cable
  • Faulty load cell
  • Scale/junction box without power

Do you need service?

If the error continues with power and cables intact, contact service for a weighing electronics or load cell fault.

Models showing this code: PR 5410, PR5410/00, PR5410/01, PR5410/03, X3

Source: Minebea Intec official user manual