No values from scale
At the internal weighing point, no value can be read from the A/D converter. Communication with the xBPI scale or digital load cells has been lost.
Check it yourself first. If it continues, it needs service.
What should you do?
Try the steps in order.
- Check that the connected scale or junction box is powered.
- Visually inspect the cables.
- Switch the device off and on again.
- Do not open the device housing. Only technical service should work on internal settings and boards.
Possible causes
- Broken cable
- Faulty load cell
- Scale/junction box without power
Do you need service?
If the error continues with power and cables intact, contact service for a weighing electronics or load cell fault.
Models showing this code: PR 5410, PR5410/00, PR5410/01, PR5410/03, X3
Source: Minebea Intec official user manual
Legal notice
The error code explanations and checks on this page are based on manufacturer manuals and are provided for general information only; they do not replace technical service, diagnosis or warranty. The meaning of a code may vary by model, firmware version and configuration.
The decision to carry out any suggested step, and its consequences, rest with the user. Terazibul cannot be held liable for device damage, weighing errors, data loss, commercial loss or any other direct or indirect damage arising from the use of this information.
On legal-for-trade (verified) scales, adjustment and calibration must only be performed by a qualified technical service in line with legal metrology rules. Disconnect power before opening the device and contact a service provider if in doubt. Brand and model names belong to their respective owners and are used only to identify the device.