Wrong serial number
The scale or load cell may have been replaced or a different scale may have been connected. The match is important for legal use.
This fault needs a technician.
What should you do?
You can safely do these while waiting for service.
- Check that the correct scale is connected to the device.
- On devices in legal (approved) use, do not break the seal. Sealed settings must be changed only by technical service.
Possible causes
- Scale or load cell replaced
- Wrong scale connected
Do you need service?
Serial number matching is tied to legal settings, so it must be done by technical service.
Models showing this code: PR 5410, PR5410/00, PR5410/01, PR5410/03, X3
Source: Minebea Intec official user manual
Legal notice
The error code explanations and checks on this page are based on manufacturer manuals and are provided for general information only; they do not replace technical service, diagnosis or warranty. The meaning of a code may vary by model, firmware version and configuration.
The decision to carry out any suggested step, and its consequences, rest with the user. Terazibul cannot be held liable for device damage, weighing errors, data loss, commercial loss or any other direct or indirect damage arising from the use of this information.
On legal-for-trade (verified) scales, adjustment and calibration must only be performed by a qualified technical service in line with legal metrology rules. Disconnect power before opening the device and contact a service provider if in doubt. Brand and model names belong to their respective owners and are used only to identify the device.