Code on display Err 75 Minebea Intec X3 PR 5410

Weight below dead load

The dead load defined at calibration is greater than the scale's current empty value, so the calibration is inconsistent.

This fault needs a technician.

What should you do?

  • On devices in legal (approved) use, do not break the seal. Sealed settings must be changed only by technical service.

Possible causes

  • Dead load determined incorrectly
  • Part removed from the structure after the dead load was taken

Do you need service?

Re-determining the dead load must be done by technical service as part of calibration.

Models showing this code: PR 5410, PR5410/00, PR5410/01, PR5410/03, X3

Source: Minebea Intec official user manual