Code on display NULL / underload display KERN & Sohn PFB

Weighing pan missing / underload

The measured value is below the weighing range. It is mostly seen when the weighing pan is not installed or is seated incorrectly.

Check it yourself first. If it continues, it needs service.

What should you do?

Try the steps in order.

  1. Install the weighing pan and check that it is seated correctly.
  2. Check that the transport lock has been removed.
  3. Turn the device off and on again after 30 seconds.

Possible causes

  • The weighing pan is not installed
  • The weighing pan is seated incorrectly
  • Transport lock not removed

Do you need service?

If the error persists with the weighing pan installed correctly, service is needed to check the load cell.

Models showing this code: PFB

Source: KERN & Sohn official user manual