Code on display Err 6 KERN & Sohn PFB

A/D converter range error

The manual describes Err 6 as an electronic error. A transport lock that has not been removed can also cause this error. The measurement signal is outside the valid range.

Check it yourself first. If it continues, it needs service.

What should you do?

Try the steps in order.

  1. Check that the transport lock has been removed.
  2. Check that the weighing pan is installed correctly.
  3. Turn the device off and on again after 30 seconds.
  • Do not open the device.

Possible causes

  • Transport lock not removed
  • The weighing pan is not installed
  • Electronics or load cell fault

Do you need service?

If the error persists with the lock removed and the balance restarted, contact the dealer/service according to the manual.

Models showing this code: PFB

Source: KERN & Sohn official user manual · Open manufacturer document