Code on display Err 3 KERN & Sohn PFB

Calibration error

According to the manual, Err 3 indicates that the wrong adjustment weight was placed during adjustment (calibration). The adjustment is not completed and the balance continues to operate with the previous adjustment.

Check it yourself first. If it continues, it needs service.

What should you do?

Try the steps in order.

  1. Check the adjustment weight value in the technical data section of the manual.
  2. Repeat the adjustment by placing the correct weight.

Possible causes

  • Calibration weight of the wrong value
  • Unstable environment during adjustment

Do you need service?

If adjustment cannot be done even with the correct weight, a service check is needed.

Models showing this code: PFB

Source: KERN & Sohn official user manual · Open manufacturer document