Zero point cannot be ınitialized
The balance cannot establish the zero point at power-on. The manual links this to a load/contact on the pan, a transport lock that was not removed, an overloaded load cell, or a main board fault, and recommends adjustment and linearization as the remedy.
Check it yourself first. If it continues, it needs service.
What should you do?
Try the steps in order.
- Empty the pan and make sure nothing touches it.
- Check that the transport lock has been removed.
- Turn the device off and on again after 30 seconds.
- Do not open the device.
Possible causes
- Load or contact on the pan
- Transport lock not removed
- Overloaded / faulty load cell
- Main board fault
Do you need service?
If the error persists after the simple checks, service is required for adjustment and linearization or a load cell check.
Models showing this code: PFB
Source: KERN & Sohn official user manual · Open manufacturer document
Legal notice
The error code explanations and checks on this page are based on manufacturer manuals and are provided for general information only; they do not replace technical service, diagnosis or warranty. The meaning of a code may vary by model, firmware version and configuration.
The decision to carry out any suggested step, and its consequences, rest with the user. Terazibul cannot be held liable for device damage, weighing errors, data loss, commercial loss or any other direct or indirect damage arising from the use of this information.
On legal-for-trade (verified) scales, adjustment and calibration must only be performed by a qualified technical service in line with legal metrology rules. Disconnect power before opening the device and contact a service provider if in doubt. Brand and model names belong to their respective owners and are used only to identify the device.