Code on display --lo-- KERN & Sohn HFD

Underload

The measured value is below zero (negative weight). It is usually seen when the hanging accessory is removed after zeroing.

Check it yourself first. If it continues, it needs service.

What should you do?

Try the steps in order.

  1. Check that the hanging accessories are installed.
  2. Turn the device off and on again after 30 seconds.
  3. If the error persists, perform an adjustment (calibration).

Possible causes

  • Removing the hanging accessory after zeroing
  • Incorrect zero point
  • Adjustment drift

Do you need service?

If the negative value continues after restarting and adjusting, service is needed to check the load cell.

Models showing this code: HFD

Source: KERN & Sohn official user manual · Open manufacturer document