Code on display Err 6 KERN & Sohn HFD

A/D converter range error

The signal from the load cell is outside the range the analog/digital converter can measure. The manual associates this with a damaged load cell or an electronics fault.

This fault needs a technician.

What should you do?

You can safely do these while waiting for service.

  1. Lower the load from the hook and turn the device off, then turn it on again after 30 seconds. If the error persists, contact service.
  • Do not open the device.
  • Do not lift loads with the scale until the fault is fixed.

Possible causes

  • Damaged load cell
  • Electronics failure
  • Damage after overload

Do you need service?

A load cell or electronic board fault is likely, so measurement and repair must be done by technical service.

Models showing this code: HFD