Zeroing range exceeded

During zeroing, the load on the hook is outside the range the device can zero. In this case, the scale cannot set the zero point correctly and weighings may be inaccurate.

Check it yourself first. If it continues, it needs service.

What should you do?

Try the steps in order.

  1. Lower the load and any hanging accessories from the hook completely.
  2. Try zeroing again with no load.
  3. Turn the device off and on again after 30 seconds.

Possible causes

  • Load or hanging accessory left on the hook
  • Preload exceeding the zeroing range
  • Load cell problem

Do you need service?

If zeroing cannot be done even with an empty hook, service is needed to check the load cell or adjustment.

Models showing this code: HFD