Zeroing range exceeded
During zeroing, the load on the hook is outside the range the device can zero. In this case, the scale cannot set the zero point correctly and weighings may be inaccurate.
Check it yourself first. If it continues, it needs service.
What should you do?
Try the steps in order.
- Lower the load and any hanging accessories from the hook completely.
- Try zeroing again with no load.
- Turn the device off and on again after 30 seconds.
Possible causes
- Load or hanging accessory left on the hook
- Preload exceeding the zeroing range
- Load cell problem
Do you need service?
If zeroing cannot be done even with an empty hook, service is needed to check the load cell or adjustment.
Models showing this code: HFD
Source: KERN & Sohn official user manual · Open manufacturer document
Legal notice
The error code explanations and checks on this page are based on manufacturer manuals and are provided for general information only; they do not replace technical service, diagnosis or warranty. The meaning of a code may vary by model, firmware version and configuration.
The decision to carry out any suggested step, and its consequences, rest with the user. Terazibul cannot be held liable for device damage, weighing errors, data loss, commercial loss or any other direct or indirect damage arising from the use of this information.
On legal-for-trade (verified) scales, adjustment and calibration must only be performed by a qualified technical service in line with legal metrology rules. Disconnect power before opening the device and contact a service provider if in doubt. Brand and model names belong to their respective owners and are used only to identify the device.