Code on display Fail KERN & Sohn HFD

Calibration error

The adjustment (calibration) failed. The manual associates the "Fail" message with a wrong adjustment weight or an incorrect adjustment procedure. The previous adjustment remains valid.

Check it yourself first. If it continues, it needs service.

What should you do?

Try the steps in order.

  1. Compare the adjustment weight value with the technical data section of the manual.
  2. Repeat the adjustment procedure from the beginning with the load still.

Possible causes

  • Adjustment weight of the wrong value
  • Load swinging during adjustment
  • Adjustment steps performed incorrectly

Do you need service?

If the error still occurs when the adjustment is repeated with the correct weight, service is needed to check the load cell.

Models showing this code: HFD

Source: KERN & Sohn official user manual · Open manufacturer document