Calibration error
The adjustment (calibration) failed. The manual associates the "Fail h" message with a wrong adjustment weight or an incorrect adjustment procedure. The previous adjustment remains valid.
Check it yourself first. If it continues, it needs service.
What should you do?
Try the steps in order.
- Compare the adjustment weight value with the technical data section of the manual.
- Repeat the adjustment procedure from the beginning with the load still.
Possible causes
- Adjustment weight of the wrong value
- Load swinging during adjustment
- Adjustment steps performed incorrectly
Do you need service?
If the error still occurs when the adjustment is repeated with the correct weight, service is needed to check the load cell.
Models showing this code: HFD
Source: KERN & Sohn official user manual · Open manufacturer document
Legal notice
The error code explanations and checks on this page are based on manufacturer manuals and are provided for general information only; they do not replace technical service, diagnosis or warranty. The meaning of a code may vary by model, firmware version and configuration.
The decision to carry out any suggested step, and its consequences, rest with the user. Terazibul cannot be held liable for device damage, weighing errors, data loss, commercial loss or any other direct or indirect damage arising from the use of this information.
On legal-for-trade (verified) scales, adjustment and calibration must only be performed by a qualified technical service in line with legal metrology rules. Disconnect power before opening the device and contact a service provider if in doubt. Brand and model names belong to their respective owners and are used only to identify the device.