Zero point cannot be ınitialized
The device cannot establish the zero point at startup. The manual associates this with a load or contact on the hook, an overloaded or faulty load cell, or a need for adjustment.
Check it yourself first. If it continues, it needs service.
What should you do?
Try the steps in order.
- Lower all the load from the hook and make sure the hook does not touch anything.
- Turn the device off and on again after 30 seconds.
- If the error persists, perform an adjustment (calibration).
- Do not open the device.
Possible causes
- Load or contact on the hook
- Overloaded / faulty load cell
- Main board fault
- Need for adjustment (calibration)
Do you need service?
If the zero point cannot be set even with an empty hook, service is needed to check the load cell or main board.
Models showing this code: HFD
Source: KERN & Sohn official user manual · Open manufacturer document
Legal notice
The error code explanations and checks on this page are based on manufacturer manuals and are provided for general information only; they do not replace technical service, diagnosis or warranty. The meaning of a code may vary by model, firmware version and configuration.
The decision to carry out any suggested step, and its consequences, rest with the user. Terazibul cannot be held liable for device damage, weighing errors, data loss, commercial loss or any other direct or indirect damage arising from the use of this information.
On legal-for-trade (verified) scales, adjustment and calibration must only be performed by a qualified technical service in line with legal metrology rules. Disconnect power before opening the device and contact a service provider if in doubt. Brand and model names belong to their respective owners and are used only to identify the device.