Code on display Err 15 KERN & Sohn HFD

Gravity constant out of range

The gravitational acceleration (g) value entered is outside the allowed range. A wrong g value causes inaccurate weighing at the location of the device.

You can usually fix this yourself.

What should you do?

Try the steps in order.

  1. Check the g value entered.
  2. Enter the correct gravity value for your location again, in the format given in the manual.
  • Do not change the adjustment menu on legal-for-trade (sealed) devices.

Possible causes

  • Gravity value entered incorrectly
  • Typing error (decimal place)

Do you need service?

The error clears when the correct gravity value is entered. On legal-for-trade devices, this setting must be made by technical service.

Models showing this code: HFD

Source: KERN & Sohn official user manual · Open manufacturer document