Code on display CAL E4 KERN & Sohn ABT

Deviation during adjustment

The balance takes several readings during adjustment and expects them to agree. If they differ too much, the adjustment is rejected. The manual gives no separate remedy for this code, so check ambient conditions first.

Check it yourself first. If it continues, it needs service.

What should you do?

Try the steps in order.

  1. Close the draft shield doors.
  2. Eliminate sources of vibration and drafts.
  3. Restart the adjustment.
  • Do not touch the balance during adjustment.

Possible causes

  • Vibration or air drafts
  • Touching the weighing pan during adjustment
  • Sudden temperature change

Do you need service?

If CAL E4 repeats in calm conditions, the weighing system must be checked by a technician.

Models showing this code: ABT 100-5NM, ABT 120-4NM, ABT 220-4NM, ABT 320-4NM, ABT 120-5DM, ABT 220-5DM, ABT 120-5DNM, ABT 220-5DNM

Source: KERN & Sohn official user manual · Open manufacturer document