Deviation during adjustment
The balance takes several readings during adjustment and expects them to agree. If they differ too much, the adjustment is rejected. The manual gives no separate remedy for this code, so check ambient conditions first.
Check it yourself first. If it continues, it needs service.
What should you do?
Try the steps in order.
- Close the draft shield doors.
- Eliminate sources of vibration and drafts.
- Restart the adjustment.
- Do not touch the balance during adjustment.
Possible causes
- Vibration or air drafts
- Touching the weighing pan during adjustment
- Sudden temperature change
Do you need service?
If CAL E4 repeats in calm conditions, the weighing system must be checked by a technician.
Models showing this code: ABT 100-5NM, ABT 120-4NM, ABT 220-4NM, ABT 320-4NM, ABT 120-5DM, ABT 220-5DM, ABT 120-5DNM, ABT 220-5DNM
Source: KERN & Sohn official user manual · Open manufacturer document
Legal notice
The error code explanations and checks on this page are based on manufacturer manuals and are provided for general information only; they do not replace technical service, diagnosis or warranty. The meaning of a code may vary by model, firmware version and configuration.
The decision to carry out any suggested step, and its consequences, rest with the user. Terazibul cannot be held liable for device damage, weighing errors, data loss, commercial loss or any other direct or indirect damage arising from the use of this information.
On legal-for-trade (verified) scales, adjustment and calibration must only be performed by a qualified technical service in line with legal metrology rules. Disconnect power before opening the device and contact a service provider if in doubt. Brand and model names belong to their respective owners and are used only to identify the device.