Code on display CAL E3 KERN & Sohn ABT

Deviation in external adjustment

If the weight placed during PCAL does not match the expected value, the balance rejects the adjustment. Usually the wrong weight was used.

Check it yourself first. If it continues, it needs service.

What should you do?

Try the steps in order.

  1. Use an adjustment weight of the value requested on the display.
  2. Place the weight in the centre of the pan.
  3. Restart the adjustment.

Possible causes

  • Adjustment weight of the wrong value
  • Dirty or damaged weight
  • Weight placed at the wrong step

Do you need service?

If CAL E3 remains with the correct weight, the balance adjustment must be checked by a technician.

Models showing this code: ABT 100-5NM, ABT 120-4NM, ABT 220-4NM, ABT 320-4NM, ABT 120-5DM, ABT 220-5DM, ABT 120-5DNM, ABT 220-5DNM