Code on display Err 24 KERN & Sohn ABT

Power voltage error

If the voltage from the adapter is outside the expected range, the balance shows this message. Correct weighing is not possible until the voltage is right.

Check it yourself first. If it continues, it needs service.

What should you do?

Try the steps in order.

  1. Check that the balance's original adapter is used.
  2. Plug the adapter firmly into the socket and the balance.
  3. Try the balance on a different outlet.

Possible causes

  • Wrong or faulty adapter
  • Loose adapter connection
  • Mains voltage problem

Do you need service?

If Err 24 remains with the correct adapter and a working socket, the power circuit must be checked by a technician.

Models showing this code: ABT 100-5NM, ABT 120-4NM, ABT 220-4NM, ABT 320-4NM, ABT 120-5DM, ABT 220-5DM, ABT 120-5DNM, ABT 220-5DNM