Unstable weight in calibration
A point taken with an unstable signal is rejected because it would lead to a wrong calibration.
This fault needs a technician.
What should you do?
- The calibration and SET-UP menu is for qualified technicians. Do not change sealed settings on legal-for-trade (approved) devices.
Possible causes
- Vibration
- Silo/tank moving or filling still in progress
Do you need service?
The calibration must be repeated by a technician under stable conditions.
Models showing this code: DGT4, DGT4AN, DGT4PB, DGT4PRONET, DGT4ETHIP, DGT4MODTCP, DGT4ETHCAT, DGT4DEVNET, DGT4CANOP
Source: Dini Argeo official user manual
Legal notice
The error code explanations and checks on this page are based on manufacturer manuals and are provided for general information only; they do not replace technical service, diagnosis or warranty. The meaning of a code may vary by model, firmware version and configuration.
The decision to carry out any suggested step, and its consequences, rest with the user. Terazibul cannot be held liable for device damage, weighing errors, data loss, commercial loss or any other direct or indirect damage arising from the use of this information.
On legal-for-trade (verified) scales, adjustment and calibration must only be performed by a qualified technical service in line with legal metrology rules. Disconnect power before opening the device and contact a service provider if in doubt. Brand and model names belong to their respective owners and are used only to identify the device.


