Dini Argeo

DGT4 error codes

21 error codes

Review the 21 error and warning messages from the Dini Argeo DGT4 manufacturer manual, with meanings and fix steps. Device type: weighing transmitter.

Covered models: DGT4, DGT4AN, DGT4PB, DGT4PRONET, DGT4ETHIP, DGT4MODTCP, DGT4ETHCAT, DGT4DEVNET, DGT4CANOP

preC Number of points not confirmed An attempt was made to take a calibration point without confirming the number of points (n tp). er.mot Unstable weight in calibration The weight was unstable when the calibration point was taken. erpnt Zero signal in calibration An empty value was read from the converter when taking the calibration point. err.x.1 Channel capacity/calibration error The capacity of channel X is not set, or there is an error in the calibration parameters. er 11 Calibration weight too low A sample weight that is too low was used in calibration. er 12 Calibration point equal to zero The calibration point taken (tp 1/tp 2/tp 3) is the same as the zero point (tp0). er 37 Balance must be calibrated The balance/channel is not calibrated. er 39 Calibration required The balance/channel is not calibrated. The weighing value is not valid. C.er. -36 Negative signal in calibration Negative internal points were calculated during calibration. C.er. -37 Calibration point insufficient Internal points below the minimum value were calculated in calibration. hw.err Hardware mismatch The software is not compatible with the installed hardware. al.err Alibi memory card missing Alibi memory is enabled but no card was detected at startup. busy Printer port busy Printing is in progress, or the transmitter is waiting to send output to the PC. unstab Printing with unstable weight Printing was attempted while the weight was unstable. un.over Printing out of range Printing was attempted while the weight was in an underload or overload condition. ------ Overload / underload Dashes on the display: the weight is above or below the measuring range. Gros.er Gross weight not positive Printing was attempted while the gross weight was zero or negative. low Below minimum weight The net weight is below the minimum value required for printing. no.0.uns No pass through zero A new printout was requested before the weight passed through net zero or became unstable. Conv Printing in unit conversion Printing was attempted while the transmitter was in unit conversion mode. err.CLk Clock board communication error The transmitter cannot communicate with the clock board.