Code on display er 11 Dini Argeo DGT4

Calibration weight too low

The manual recommends using a weight of at least half of capacity.

This fault needs a technician.

What should you do?

  • The calibration and SET-UP menu is for qualified technicians. Do not change sealed settings on legal-for-trade (approved) devices.

Possible causes

  • Insufficient calibration weight

Do you need service?

The calibration must be repeated with a suitable weight.

Models showing this code: DGT4, DGT4AN, DGT4PB, DGT4PRONET, DGT4ETHIP, DGT4MODTCP, DGT4ETHCAT, DGT4DEVNET, DGT4CANOP

Source: Dini Argeo official user manual