Message on display - OUT CAPACITY ! - / UNDERLOAD Dini Argeo 3590E

Underload

This occurs when the reading falls below gross zero by 100 divisions on an approved device, or below capacity + 9 divisions on a non-approved device. It usually means the platform was lifted or the zero has drifted.

Check it yourself first. If it continues, it needs service.

What should you do?

Try the steps in order.

  1. Check that the platform is seated properly and is not jammed.
  2. Zero with the ZERO key on an empty platform.
  3. Switch the device off and on again.

Possible causes

  • Platform or pan dislodged
  • Removal of a container that was zeroed with tare
  • Zero point drift

Do you need service?

If the warning persists on an empty, properly seated platform, the load cell or cable must be checked by service.

Models showing this code: 3590EKR, 3590EXP, 3590EXT, 3590EDX, CPWE, CPWET

Source: Dini Argeo official user manual