Range 1 value is zero
Because the first range capacity is 0, the indicator cannot calculate weight. The technical default setting must be loaded and the calibration redone.
This fault needs a technician.
What should you do?
- The calibration and SET-UP menu is for qualified technicians. Do not change sealed settings on legal-for-trade (approved) devices.
Possible causes
- Capacity parameter not entered
- Corrupted setup data
Do you need service?
The capacity must be entered and the calibration renewed. This requires a technician.
Models showing this code: 3590EKR, 3590EXP, 3590EXT, 3590EDX, CPWE, CPWET
Source: Dini Argeo official user manual
Legal notice
The error code explanations and checks on this page are based on manufacturer manuals and are provided for general information only; they do not replace technical service, diagnosis or warranty. The meaning of a code may vary by model, firmware version and configuration.
The decision to carry out any suggested step, and its consequences, rest with the user. Terazibul cannot be held liable for device damage, weighing errors, data loss, commercial loss or any other direct or indirect damage arising from the use of this information.
On legal-for-trade (verified) scales, adjustment and calibration must only be performed by a qualified technical service in line with legal metrology rules. Disconnect power before opening the device and contact a service provider if in doubt. Brand and model names belong to their respective owners and are used only to identify the device.