Span too high
During weight adjustment, the signal from the load cell is higher than expected. There may be an abnormality in the load cell. In count/percent mode it also appears when the sample weight is too high.
This fault needs a technician.
What should you do?
You can safely do these while waiting for service.
- If it appeared in count or percent mode, repeat sampling with a lighter sample.
- Do not allow unauthorized persons to use calibration (weight adjustment) mode.
Possible causes
- High load cell output
- Load cell fault
- Sample too heavy (in count/percent mode)
Do you need service?
If it appears during weight adjustment, readjustment from the CAL 4 zeroing step and a load cell check must be done by service.
Models showing this code: CPS-PLUS
Source: CAS official user manual
Legal notice
The error code explanations and checks on this page are based on manufacturer manuals and are provided for general information only; they do not replace technical service, diagnosis or warranty. The meaning of a code may vary by model, firmware version and configuration.
The decision to carry out any suggested step, and its consequences, rest with the user. Terazibul cannot be held liable for device damage, weighing errors, data loss, commercial loss or any other direct or indirect damage arising from the use of this information.
On legal-for-trade (verified) scales, adjustment and calibration must only be performed by a qualified technical service in line with legal metrology rules. Disconnect power before opening the device and contact a service provider if in doubt. Brand and model names belong to their respective owners and are used only to identify the device.