Code on display Err 09 CAS CPS-PLUS

Out of zero range

Zeroing failed because the current weight is outside the range in which the zero key can operate. The zero key works only within a small percentage of capacity.

You can usually fix this yourself.

What should you do?

Try the steps in order.

  1. Remove the load from the platform.
  2. Check that the platform is free and not touching anything.
  3. Press the zero key again. Use the tare key instead of zeroing when there is a load.
  • Do not set the zero range (F13) to high values. The manual warns that 10% or more can damage the load cell.

Possible causes

  • Zeroing attempted with a load on the platform
  • Platform touching something

Do you need service?

If zeroing is still not possible with an empty platform, the zero point has shifted and service must check the calibration.

Models showing this code: CPS-PLUS

Source: CAS official user manual