Code on display Err 01 CAS CPS-PLUS

Initial stability error

Startup could not be completed because the weight was swinging (unstable) while the scale was making its initial zero setting at power-on. According to the manual, startup cannot complete while the weight is vibrating.

You can usually fix this yourself.

What should you do?

Try the steps in order.

  1. Place the scale on a level, vibration-free surface.
  2. Switch the balance off and on again.
  • Do not touch the platform or leave any load on it during power-on.

Possible causes

  • Vibrating or sloping floor
  • Moving load on the platform at power-on
  • Air draft or wind

Do you need service?

If the error persists when the scale is turned on again on a suitable surface, the platform and load cell connection must be checked by service.

Models showing this code: CPS-PLUS

Source: CAS official user manual