Code on display Err 13 CAS CPS-PLUS

Zero adjustment value out of range

The zero point value determined during weight adjustment (calibration) is outside the accepted range. Calibration cannot be completed unless the zero point is taken correctly.

Check it yourself first. If it continues, it needs service.

What should you do?

Try the steps in order.

  1. Check that the platform is empty and not touching anything.
  2. Switch the balance off and on again.

Possible causes

  • Load left on the platform during adjustment
  • Platform touching something
  • Load cell fault

Do you need service?

If the zero point is still out of range with an empty platform, service must check the load cell and repeat the weight adjustment.

Models showing this code: CPS-PLUS

Source: CAS official user manual