Balance not level
Readings are wrong when the balance is tilted. The bubble in the level indicator must be centered.
You can usually fix this yourself.
What should you do?
Try the steps in order.
- Check the bubble in the level indicator.
- Turn the feet until the bubble is inside the circle.
- Calibrate the balance after leveling.
Possible causes
- Balance moved
- Sloping bench
- The feet adjustment was disturbed
Do you need service?
The warning clears once the balance is level.
Models showing this code: BCA64-1S, BCA623-1S
Source: Sartorius official user manual
Legal notice
The error code explanations and checks on this page are based on manufacturer manuals and are provided for general information only; they do not replace technical service, diagnosis or warranty. The meaning of a code may vary by model, firmware version and configuration.
The decision to carry out any suggested step, and its consequences, rest with the user. Terazibul cannot be held liable for device damage, weighing errors, data loss, commercial loss or any other direct or indirect damage arising from the use of this information.
On legal-for-trade (verified) scales, adjustment and calibration must only be performed by a qualified technical service in line with legal metrology rules. Disconnect power before opening the device and contact a service provider if in doubt. Brand and model names belong to their respective owners and are used only to identify the device.