Code on display APP.ERR. Sartorius Entris II Advanced BCA

Invalid weight for application

The running application cannot calculate a result with this value. The load is below the minimum, negative, or there is no sample.

You can usually fix this yourself.

What should you do?

Try the steps in order.

  1. Place the sample on the pan.
  2. Increase the load above the minimum.
  3. If the value is negative, empty the pan and zero the balance.

Possible causes

  • Sample below minimum load
  • Negative weight value
  • No sample on the pan

Do you need service?

The warning clears once a suitable sample is placed.

Models showing this code: BCA64-1S, BCA623-1S

Source: Sartorius official user manual