Code on display ERR 54 Sartorius Entris II Advanced BCA

Weighing signal too low

The balance detects a missing pan, or a load removed after power-on, this way. If the pan is in place, the weighing system or electronics may be faulty.

Check it yourself first. If it continues, it needs service.

What should you do?

Try the steps in order.

  1. Install the weighing pan.
  2. Switch the balance off and on again.
  3. Make sure the pan is empty at power-on.

Possible causes

  • Weighing pan not installed
  • Forgotten weight removed after power-on
  • Weighing system or electronics failure

Do you need service?

If the warning remains with the pan fitted and an empty start, the weighing system must be checked by a technician.

Models showing this code: BCA64-1S, BCA623-1S

Source: Sartorius official user manual