Calibration canceled
The balance could not start calibration because of a zero point error or because it is not level. A load on the pan or strong drift also causes this.
Check it yourself first. If it continues, it needs service.
What should you do?
Try the steps in order.
- Empty the weighing pan.
- Zero the balance.
- Wait for the warm-up time.
- Level the balance using the level indicator.
- Restart the calibration.
Possible causes
- Balance not zeroed before calibration
- Sample left on the pan
- Strong drift before warm-up ends
- Balance not level (verified models)
Do you need service?
If it is still canceled after these steps, the internal calibration weight mechanism must be checked by a technician.
Models showing this code: BCA64-1S, BCA623-1S
Source: Sartorius official user manual
Legal notice
The error code explanations and checks on this page are based on manufacturer manuals and are provided for general information only; they do not replace technical service, diagnosis or warranty. The meaning of a code may vary by model, firmware version and configuration.
The decision to carry out any suggested step, and its consequences, rest with the user. Terazibul cannot be held liable for device damage, weighing errors, data loss, commercial loss or any other direct or indirect damage arising from the use of this information.
On legal-for-trade (verified) scales, adjustment and calibration must only be performed by a qualified technical service in line with legal metrology rules. Disconnect power before opening the device and contact a service provider if in doubt. Brand and model names belong to their respective owners and are used only to identify the device.