Code on display Err9 RADWAG WLC

Internal weight mechanism timeout

The motorized lifting or lowering of the internal adjustment weight was not completed in the expected time. This code appears only on WLC models with an internal adjustment weight (C/2). A mechanism fault prevents automatic calibration.

This fault needs a technician.

What should you do?

You can safely do these while waiting for service.

  1. Turn the balance off and on again after 30 seconds. If the error persists, contact service.
  • Do not open the device.
  • Until service arrives, adjustment can be made with an external adjustment weight.

Possible causes

  • Internal adjustment weight mechanism jammed
  • Motor or control fault

Do you need service?

The internal adjustment weight mechanism is inside the device and can only be repaired by technical service.

Models showing this code: WLC

Source: RADWAG official user manual