Code on display Err2 RADWAG WLC

Zeroing range error

The balance cannot be zeroed because, when the zero key is pressed, the indication is outside the permitted zeroing range (±2% of maximum capacity). This limit prevents an actual load from being zeroed by mistake.

You can usually fix this yourself.

What should you do?

Try the steps in order.

  1. Empty the weighing pan completely.
  2. Press the zero key again once the indication is stable.
  3. If you are using a container, tare instead of zeroing.

Possible causes

  • A load or container is on the weighing pan
  • An object was placed on the weighing pan after switch-on
  • The weighing pan is dirty or installed incorrectly

Do you need service?

Zeroing can be done once the weighing pan is emptied. If it persists on an empty pan, a service check is required.

Models showing this code: WLC