Measuring range exceeded
The load is above the upper limit the balance can measure, so no value can be displayed. On precision balances, an overload can damage the measuring system.
Check it yourself first. If it continues, it needs service.
What should you do?
Try the steps in order.
- Remove the load from the weighing pan immediately.
- Weigh the sample in portions that fit the capacity.
- Do not place a load above capacity or an impact load on the pan.
Possible causes
- Sample heavier than capacity
- Pressure or impact on the weighing pan
Do you need service?
If FULL is also shown with an empty pan, the measuring system must be checked by technical service.
Source: RADWAG official user manual · Open manufacturer document
Legal notice
The error code explanations and checks on this page are based on manufacturer manuals and are provided for general information only; they do not replace technical service, diagnosis or warranty. The meaning of a code may vary by model, firmware version and configuration.
The decision to carry out any suggested step, and its consequences, rest with the user. Terazibul cannot be held liable for device damage, weighing errors, data loss, commercial loss or any other direct or indirect damage arising from the use of this information.
On legal-for-trade (verified) scales, adjustment and calibration must only be performed by a qualified technical service in line with legal metrology rules. Disconnect power before opening the device and contact a service provider if in doubt. Brand and model names belong to their respective owners and are used only to identify the device.