Code on display -Err2- RADWAG PS R2

Outside zeroing range

According to the manual, zeroing can only be done within ±2% of maximum capacity. When the display is outside this limit, the balance refuses to zero. This limit prevents an actual load on the pan from being zeroed by mistake.

Check it yourself first. If it continues, it needs service.

What should you do?

Try the steps in order.

  1. Empty the weighing pan completely.
  2. Check that the weighing pan is seated correctly.
  3. Press the zero key again once the indication is stable.
  4. If you are using a container, tare instead of zeroing.

Possible causes

  • Sample or container on the pan
  • Pan installed incorrectly
  • An object touching the pan

Do you need service?

If Err2 continues with the pan empty and correctly installed, the balance's zero point must be checked by technical service.

Source: RADWAG official user manual · Open manufacturer document