Tarring out of range
Taring is not possible at this value, for example when the reading is negative. The message suggests pressing the zero key or restarting.
You can usually fix this yourself.
What should you do?
Try the steps in order.
- Empty the pan and press the zero key.
- Place the container on the pan and tare again once the indication is stable.
- Restart the analyzer if the problem continues.
Possible causes
- Tare taken with a negative display
- Tare attempted after the load was removed from the pan
Do you need service?
The message clears once taring is repeated after zeroing.
Source: RADWAG official user manual · Open manufacturer document
Legal notice
The error code explanations and checks on this page are based on manufacturer manuals and are provided for general information only; they do not replace technical service, diagnosis or warranty. The meaning of a code may vary by model, firmware version and configuration.
The decision to carry out any suggested step, and its consequences, rest with the user. Terazibul cannot be held liable for device damage, weighing errors, data loss, commercial loss or any other direct or indirect damage arising from the use of this information.
On legal-for-trade (verified) scales, adjustment and calibration must only be performed by a qualified technical service in line with legal metrology rules. Disconnect power before opening the device and contact a service provider if in doubt. Brand and model names belong to their respective owners and are used only to identify the device.